Cation Analysis - CAT1™ Column

 

  • High Resolution

  • Use Conventional HPLC Systems


Transitional Metals
Mono and Divalent Cations
Ask Technical Questions

CAT1 Cation Analysis Column
The Sarasep CAT1 column permits high resolution separation of alkali, alkaline earth, transition or heavy metals using conventional HPLC equipment with UV or conductivity detection.

Simultaneous determination of alkali and alkaline earth metals is possible.

The versatility of the Sarasep CAT1 column is demonstrated by simply changing the eluent from 0.1M Ce(III) to ethylenediamine citrate, thus allowing determination of transition and heavy metals.

CAT1

Guard
Column
$ 220
Guard
Cartridges
$ 120 (2/pkg)
250 x 4.6mm
$ 700
   CAT1 Column 5093-G 5093-CP 5093

Guard cartridges require PEEK Guard holder, P/N 5053 @ $ 90.

Transition and Heavy Metals Analysis

Column: CAT1 Cation Column
Order No: 5093
Eluent: 3.5mM ethylenediamine
10mM citric acid
Gradient: None
Flow Rate: 1.0 ml/min
Temp: ambient
Detection: conductivity
Sample: 1. Cu++
  2. Ni++
  3. Zn++
  4. Fe++
  5. Mn++
  6. Cd++
  7. Pb++

Mono and Divalent Cations

Column: CAT1 Cation Column
Order No: 5093
Eluent: 3.5mM ethylenediamine
10mM citric acid
Gradient: None
Flow Rate: 1.0 ml/min
Temp: ambient
Detection: conductivity
Sample: 1. Na+
  2. NH4+
  3. K+
  4. Rb+
  5. Cs+
  6. Mg++
  7. Ca++

 

   HOME

This site is developed and maintained by Isoft Solutions
www.isoftindia.com

Best viewed in Microsoft Internet Explorer 4.0 or with 800X600 resolution.